CREATIVE MEDICAL FINGERTIP OXIMETER
OLED DISPLAYS
A. SPO2 MEASUREMENT
SENSOR: DUAL-WAVELENGTH LED SENSOR WITH WAVELENGTH: RED LIGHT: 663 NM, INFRARED LIGHT: 890 NM. MAXIMAL AVERAGE OPTICAL OUTPUT POWER: ≤2MW
SPO2 DISPLAY RANGE: 35% - 100%
SPO2 MEASURING ACCURACY: ≤ 2% FOR SPO2 RANGE FROM 70% TO 100%
B. PULSE RATE MEASUREMENT
PR DISPLAY RANGE: 30 BPM – 240 BPM
PR MEASURING ACCURACY: ±2BPM OR ±2% (WHICHEVER IS GREATER)
C. PERFUSION INDEX (PI) DISPLAY RANGE
0% - 20%
D. OVER-LIMIT SETTINGS
SPO2:
LOW LIMIT SETTING RANGE: 85% - 99%, STEP: 1%
DEFAULT SETTING: 90%
PULSE RATE:
LOW LIMIT SETTING RANGE: 30 - 60 BPM, STEP: 1BPM;
HIGH LIMIT SETTING RANGE: 100 - 240 BPM, STEP: 5BPM;
DEFAULT SETTING: HIGH: 120BPM; LOW: 50BPM
E. AUDIBLE & VISUAL ALERT FUNCTION
WHEN MEASURING, IF SPO2 VALUE OR PULSE RATE VALUE EXCEEDS THE PRESET LIMIT, THE DEVICE WILL ALERT WITH BEEP AUTOMATICALLY AND THE VALUE WHICH EXCEEDS LIMIT WILL FLASH ON THE SCREEN.
F. POWER SUPPLY REQUIREMENT
2 X LR03 (AAA) ALKALINE BATTERIES SUPPLY VOLTAGE: 3.0VDC, OPERATING CURRENT: ≤40MA
G. ENVIRONMENTAL CONDITIONS
OPERATING TEMPERATURE: 5°C - 40°C
OPERATING HUMIDITY: 30% - 80%
STORAGE TEMPERATURE: -20°C - 60°C
STORAGE HUMIDITY: 10% - 95%
ATMOSPHERIC PRESSURE: 70KPA - 106KPA
H. LOW PERFUSION PERFORMANCE
THE ACCURACY OF SPO2 AND PR MEASUREMENT STILL MEET THE PRECISION DESCRIBED ABOVE WHEN THE MODULATION AMPLITUDE IS AS LOW AS 0.6%.
I. AMBIENT LIGHT INTERFERENCE
THE DIFFERENCE BETWEEN THE SPO2 VALUE MEASURED IN THE CONDITION OF INDOOR NATURAL LIGHT AND THAT OF DARKROOM IS LESS THAN ±1%.
J. DIMENSIONS: 56 MM (L) × 34 MM (W) × 30 MM (H)
NET WEIGHT: APPROX. 60G (INCLUDING BATTERIES)
K. PRODUCT ACCESSORIES:
ONE FINGERTIP OXIMETER
ONE LANYARD
ONE USER MANUAL
2 AAA ALKALINE BATTERIES
Place of Origin
China
Benefit
MEASURING THE PULSE RATE AND FUNCTIONAL OXYGEN SATURATION (SPO2) THROUGH A PATIENT’S FINGER. IT IS APPLICABLE FOR SPOT-CHECKING SPO2 AND PULSE RATE OF ADULT AND PEDIATRIC PATIENTS IN HOMES AND MEDICAL CLINICS.
SMALL AND PORTABLE
LIGHTWEIGHT BODY CAN POCKETED, CARRY IT WITH YOU, AND MEASURE AT ANY TIME
FEATURES
DISPLAY SPO2, PR, PI, AND PLETHYSMOGRAM
AUTO POWER ON/OFF CHANGE BETWEEN PR AND PI OVER-LIMIT INDICATION
LOW BATTERY INDICATION
FOUR DIRECTION DISPLAY
SETTING MENU (INCLUDING OVER-LIMIT SETTING)
THE MANUFACTURER HEREBY DECLARES THAT THIS DEVICE COMPLIES WITH THE
FOLLOWING STANDARDS: IEC 60601-1: 2005+A1: 2012 MEDICAL ELECTRICAL EQUIPMENT-PART
1: GENERAL REQUIREMENTS FOR BASIC SAFETY AND ESSENTIAL
PERFORMANCE; ISO 80601-2-61: 2017 - MEDICAL ELECTRICAL EQUIPMENT -- PART
2-61: PARTICULAR REQUIREMENTS FOR BASIC SAFETY AND ESSENTIAL
PERFORMANCE OF PULSE OXIMETER EQUIPMENT. FOLLOW THE PROVISIONS OF THE COUNCIL DIRECTIVE MDD 93/42/EEC.
EU CERTIFIED
1 YEAR WARRANTY
Product Usage
1. Install two AAA batteries into battery cassette before covering its cover.
2. Plug one finger into rubber hole of the Oximeter (it is best to plug the finger thoroughly) before releasing the clamp with the nail upwards.
3. Press button on the front panel. FOR DETAILS, PLEASE REFER TO USER MANUAL
Ingredients
Plastic
Shipping Information
No
Pieces per Container
1 piece/ box
Storage Condition
OPERATING TEMPERATURE: 5°C - 40°C
OPERATING HUMIDITY: 30% - 80%
STORAGE TEMPERATURE: -20°C - 60°C
STORAGE HUMIDITY: 10% - 95%
ATMOSPHERIC PRESSURE: 70KPA - 106KPA
Warnings
CAUTION
WHEN PLUGGING YOUR FINGER INTO THE OXIMETER
Width
3.4
Height
3
Depth
5.6
1. Home Delivery (except products prohibited by Department of Health or shipped by suppliers)
Free shipping for net order value upon $399 (except products shipped by suppliers). Express Order during 9am - 7pm will be delivered as fast as 30 mins.
2. Click & Collect (except products shipped by suppliers)
Over 160 Watsons Pick Up Points. Support Click and Collect Express in as fast as 30 mins.
3. SF Locker (except products prohibited by Department of Health or shipped by suppliers)
Free SF Locker Pick Up Points Upon Purchase of $250, located all over Hong Kong, including residential areas, estate shopping malls.